Ohio Administrative Code|Rule 3701:1-66-07 | Fluoroscopic radiation-generating equipment.

                                                

(A) Fluoroscopic equipment shall meet the
following standards:

(1) Unless the United
States food and drug administration (FDA) has granted a variance for specific
fluoroscopic equipment, the source-to-skin distance (SSD) for fluoroscopy
equipment shall not be less than:

(a) Thirty-eight centimeters on stationary fluoroscopic equipment
unless a particular procedure application prohibits that distance, in which
case the SSD shall not be less than twenty centimeters;

(b) Thirty centimeters on mobile fluoroscopic equipment unless a
particular procedure prohibits that distance, in which case it shall not be
less than twenty centimeters; and

(c) Nineteen centimeters for c-arm type fluoroscopic equipment
having a maximum source-to-image distance (SID) less than forty-five
centimeters unless a particular procedure prohibits that distance, in which
case it shall not be less than ten centimeters. Such systems shall be used for
extremity or dental purposes only;

(2) For c-arm fluoroscopic equipment
equipped with a removable spacer cone, the spacer cone shall be attached to the
x-ray source during use at all times unless it interferes with the clinical
procedure;

(3) The fluoroscopic imaging assembly
shall be provided with a primary protective barrier which intercepts the entire
cross-section of the useful beam at any SID and shall prevent further exposures
when the primary barrier is not in the path of the entire x-ray
beam;

(4) All fluoroscopic equipment shall
provide intensified imaging. As used in this rule "intensified
imaging" will include the use of digital image receptors;

(5) Fluoroscopic equipment shall meet the
following field limitation specifications:

(a) For fluoroscopic equipment manufactured before June 10, 2006,
the following applies:

(i) Neither the length
nor the width of the x-ray field in the plane of the image receptor shall
exceed that of the visible area of the image receptor by more than three per
cent of the SID. The sum of the excess length and the excess width shall be no
greater than four per cent of the SID; and

(ii) For rectangular
x-ray fields used with circular image receptors, the error in alignment shall
be determined along the length and width dimensions of the x-ray field which
pass through the center of the visible area of the image receptor;

(b) For fluoroscopic equipment with a circular image receptor
manufactured on or after June 10, 2006, the maximum area of the x-ray field in
the plane of the circular image receptor shall conform with one of the
following requirements:

(i) When any linear
dimension of the visible area of the image receptor measured through the center
of the visible area is less than or equal to thirty-four centimeters in any
direction, at least eighty per cent of the area of the x-ray field shall
overlap the visible area of the image receptor; or

(ii) When any linear
dimension of the visible area of the image receptor measured through the center
of the visible area is greater than thirty-four centimeters in any direction,
the x-ray field measured along the direction of greatest misalignment with the
visible area of the image receptor shall not extend beyond the edge of the
visible area of the image receptor by more than two centimeters;

(c) For fluoroscopic equipment with a rectangular image receptor
manufactured on or after June 10, 2006, the following applies:

(i) Neither the length
nor the width of the x-ray field in the plane of the image receptor shall
exceed that of the visible area of the image receptor by more than three per
cent of the SID. The sum of the excess length and the excess width shall be no
greater than four per cent of the SID; and

(ii) The error in
alignment shall be determined along the length and width dimensions of the
x-ray field which pass through the center of the visible area of the image
receptor;

(d) If the fluoroscopic x-ray field size is adjusted
automatically as the SID or image receptor size is changed, a capability may be
provided for overriding the automatic adjustment in case of system failure. If
it is so provided, a signal visible at the operator's position shall
indicate whenever the automatic field adjustment is overridden. Each such
system failure override switch shall be clearly labeled as
follows:

"For X-ray Field Limitation System
Failure;"

(e) Beam-limiting devices shall be provided with a means for
stepless adjustment of the x-ray field; and

(f) Stepless adjustment shall, at the greatest SID, provide
continuous field sizes from the maximum obtainable to a field size of five
centimeters by five centimeters or less;

(6) Timers shall meet the following
specifications:

(a) A means shall be provided to preset the cumulative on-time
timer of the fluoroscopic tube. The maximum cumulative time of the timer shall
not exceed five minutes without resetting;

(b) The timer shall terminate the exposure or emit a signal
audible to the operator when the exposure time reaches a maximum of five
minutes. The signal shall continue to sound while x-rays are produced until the
timer is reset;

(c) For x-ray controls manufactured on or after June 10, 2006,
there shall be provided for each fluoroscopic tube:

(i) A display of the
fluoroscopic irradiation time at the operator's working position. This
display shall function independently of the audible signal described in
paragraph (A)(6)(c)(ii) of this rule. The following requirements
apply:

(a) When the x-ray tube
is activated, the fluoroscopic irradiation time in minutes and tenths of
minutes shall be continuously displayed and updated at least once every six
seconds;

(b) The fluoroscopic
irradiation time shall also be displayed within six seconds of termination of
an exposure and remain displayed until reset; and

(c) Means shall be
provided to reset the display to zero prior to the beginning of a new
examination or procedure;

(ii) A signal audible to
the operator shall sound for each passage of five minutes of fluoroscopic
irradiation time during an examination or procedure. The signal shall sound
until manually reset or, if automatically reset, for at least two
seconds;

(7) X-ray production in the fluoroscopic
mode shall be controlled by a device which requires continuous pressure by the
operator for the entire time of any exposure. When recording serial
fluoroscopic images, the operator shall be able to terminate the x-ray exposure
at any time, but means may be provided to permit completion of any single
exposure of the series in progress;

(8) Fluoroscopic systems shall meet the
following air kerma rate limits:

(a) Fluoroscopic equipment provided with only automatic exposure
rate control, or provided with both automatic exposure rate control and manual
mode capabilities, shall not exceed an air kerma rate of eighty-eight milligray
per minute (ten roentgens per minute exposure rate) in either mode at any
combination of tube potential and current, at the measurement point specified
in paragraph (C)(6) of this rule;

(b) Fluoroscopic equipment provided with only manual mode
capabilities shall not exceed an air kerma rate of forty-four milligray per
minute (five roentgens per minute exposure rate) at any combination of tube
potential and current, at the measurement point specified in paragraph (C)(6)
of this rule; and

(c) For fluoroscopic equipment that is provided with high-level
control, and the high-level control is activated, the air kerma rate shall not
exceed one hundred seventy-six milligray per minute (twenty roentgens per
minute exposure rate) at any combination of tube potential and current, at the
measurement point specified in paragraph (C)(6) of this rule;

(i) For all fluoroscopy
equipment that is provided with high-level control, special means of activation
of high level control, such as manual pressure applied continuously by the
operator, shall be required to avoid accidental use; and

(ii) A continuous signal
audible to the operator shall indicate that high level control is being
employed;

(9) During fluoroscopy
and cinefluorography the x-ray tube potential and current shall be continuously
indicated;

(10) For undertable
fluoroscopic equipment, a shielding device of at least 0.25 millimeter lead
equivalent shall cover the bucky-slot;

(11) For undertable
fluoroscopic equipment, protective drapes, or other devices, at least 0.25
millimeter lead equivalent shall be provided between the patient and the
individual operating the fluoroscopic equipment to intercept scattered
radiation which would otherwise reach the fluoroscopist and others near the
x-ray unit, except when such drapes or other devices would compromise the
sterile field. Such devices shall not substitute for wearing required
protective apparel;

(12) Radiography using
the fluoroscopic imaging assembly shall meet the following
specifications:

(a) A means shall be provided between the source and the patient
which will automatically limit the x-ray field at the time the exposure is
initiated to no more than the portion of the image receptor selected by the
operator for spot films or radiographic images. If the x-ray field size is less
than the size of the selected portion of the image receptor, the field size
shall not open automatically to the size of the selected portion of the image
receptor unless the operator has selected such a mode of
operation;

(b) Neither the length nor the width of the x-ray field in the
plane of the image receptor shall differ from the corresponding dimensions of
the selected portion of the image receptor by more than three per cent of the
SID when adjusted for full coverage of the selected portion of the image
selector;

(c) The center of the x-ray field in the plane of the image
receptor shall be aligned with the center of the selected portion of the image
receptor to within two per cent of the SID; and

(d) Means shall be provided to reduce the x-ray field size in the
plane of the image receptor to a size smaller than the selected portion of the
image receptor. The minimum field size at the greatest SID shall not exceed
five centimeters by five centimeters;

(13) Fluoroscopic equipment manufactured
on or after June 10, 2006, shall display at the operator's working
position the air kerma rate (AKR) and cumulative air kerma in accordance with
the following requirements:

(a) When the x-ray tube is activated and the number of images
produced per unit time is greater than six images per second, the AKR in
milligrays per minute shall be continuously displayed and updated at least once
every second;

(b) The cumulative air kerma in units of milligrays shall be
displayed either within five seconds of termination of an exposure or displayed
continuously and updated at least once every five seconds;

(c) The display of the AKR shall be clearly distinguishable from
the display of the cumulative air kerma;

(d) The AKR and cumulative air kerma shall represent the value
for conditions of free-in-air irradiation at one of the following reference
locations specified according to the type of fluoroscope;

(i) For fluoroscopes with
x-ray source below the x-ray table, x-ray source above the table, or of lateral
type, the reference location shall be the respective locations specified in
paragraph (C)(6)(a), (C)(6)(b) or (C)(6)(d) of this rule; or

(ii) For C-arm
fluoroscopes, the reference location shall be fifteen centimeters from the
isocenter toward the x-ray source along the beam axis. Alternatively, the
reference location shall be at a point specified by the manufacturer to
represent the location of the intersection of the x-ray beam with the
patients skin;

(e) Means shall be provided to reset to zero the display of
cumulative air kerma prior to the commencement of a new examination or
procedure; and

(f) The displayed AKR and cumulative air kerma shall not deviate
from the actual values by more than plus or minus thirty-five per
cent;

(14) Fluoroscopic
equipment manufactured on or after June 10, 2006 shall be equipped with means
to display a last image hold (LIH) image following termination of the
fluoroscopic exposure:

(a) For a LIH image obtained by retaining pre-termination
fluoroscopic images, if the number of images and method of combining images are
selectable by the user, the selection shall be indicated prior to initiation of
the fluoroscopic exposure;

(b) For a LIH image obtained by initiating a separate
radiographic exposure at termination of the fluoroscopic imaging, the technique
factors for the LIH image shall be selectable prior to the fluoroscopic
exposure, and the combination selected shall be indicated prior to initiation
of the fluoroscopic exposure; and

(c) Means shall be provided to clearly indicate to the user
whether a displayed image is the LIH radiograph or fluoroscopy. Display of the
LIH radiograph shall be replaced by the fluoroscopic image concurrently with
re-initiation of the fluoroscopic exposure unless separate displays are
provided.

(B) In addition to other applicable radiation safety rules
adopted pursuant to Chapter 3748. of the Revised Code, handlers of fluoroscopic
radiation-generating equipment shall comply with the following:

(1) Any individual who is
in the room during the fluoroscopic procedure shall be adequately protected by
standing behind a whole body protective barrier or shall be required to wear a
protective lead apron of not less than 0.25 millimeter lead
equivalent;

(2) Protective lead or
lead equivalent gloves shall be used by individuals who are required to have
their hands in or near the useful beam; and

(3) Handlers of fluoroscopic equipment
used for interventional or cardiac procedures or on pediatric or pregnant
patients shall maintain a record of:

(a) Cumulative air kerma or dose area product used for each
examination, if the display of either is available on the fluoroscopic
equipment; or

(b) The following items if the cumulative air kerma or dose area
product is not displayed on the fluoroscopic equipment:

(i) Mode of operation
such as high-level or pulsed mode;

(ii) Cumulative
fluoroscopic exposure time; and

(iii) Number of
radiographs and number of acquisitions.

(C) In addition to other applicable quality assurance
requirements of Chapter 3701:1-66 of the Administrative Code, handlers of
fluoroscopic equipment shall comply with the following:

(1) Handlers shall
designate and utilize a radiation expert who shall develop in writing and
perform fluoroscopic image quality evaluations appropriate for the fluoroscopic
equipment including written procedures to include time intervals and system
conditions for the evaluation of image quality;

(2) On new installations
or reinstallations of existing equipment prior to patient exposure, handlers
shall utilize a radiation expert to perform the following:

(a) Radiographic device tests to determine compliance with
allowable limits as specified in paragraph (A)(12) of this rule;

(b) Fluoroscopic image quality evaluations as specified in
paragraph (C)(1) of this rule;

(c) Air kerma rate tests as specified in paragraph (C)(6) of this
rule;

(d) High contrast and low contrast resolution evaluations
in both fluoroscopic and radiographic modes;

(e) Five minute timer evaluations; and

(f) Evaluation of the accuracy of technique factor
indicators and integrated radiation dose displays;

(3) After initial
evaluations of fluoroscopic equipment have been performed, the test and
evaluations in paragraph (C)(2) of this rule shall be performed by a radiation
expert annually;

(4) After repair or
replacement of any component of the fluoroscopic equipment which may alter the
radiation output or image quality, prior to patient use, a radiation expert
shall perform and document measurements of air kerma rates as specified in
paragraph (C)(6) of this rule and image quality as specified in paragraph
(C)(1) of this rule unless in the documented determination of a radiation
expert, the repair or replacement will not cause a significant change in
radiation output or significant degradation of image quality as specified in
the quality assurance program;

(a) The radiation expert may designate qualified individuals to
perform and document the measurements specified in paragraphs (C)(6) and (C)(1)
of this rule;

(b) The radiation expert shall provide the criteria for
qualifying these designees in the quality assurance program; and

(c) The radiation expert's approval of the designee's
test results shall be documented within thirty days;

(5) The results of all tests performed in
accordance with paragraphs (C)(2) to (C)(4) of this rule shall:

(a) Include the technique factors used in determining such
results;

(b) Include the name of the individual performing the
measurements;

(c) Include the date the measurements were performed;
and

(d) Be maintained by the IRRP between inspections for review by
the department;

(6) Compliance with air
kerma rate allowable limits in paragraph (A)(8) of this rule shall be
determined as follows:

(a) If the source is below the x-ray table, the air kerma rate
shall be measured at one centimeter above the tabletop or cradle;

(b) If the source is above the x-ray table, the air kerma rate
shall be measured at thirty centimeters above the tabletop with the end of the
beam limiting device or spacer positioned as closely as possible to the point
of measurement;

(c) For c-arm type fluoroscopic equipment, the air kerma rate
shall be measured at thirty centimeters from the input surface of the image
receptor with the source positioned at any SID;

(d) For fixed SID lateral fluoroscopes attached to the x-ray
table, the maximum air kerma rate shall be measured at a point fifteen
centimeters from the centerline of the x-ray table and in the direction of the
x-ray source with the end of the beam-limiting device or spacer positioned as
closely as possible to the point of measurement. If the table top is moveable,
it shall be positioned as closely as possible to the lateral x-ray source, with
the end of the beam-limiting device or spacer no closer than fifteen
centimeters to the centerline of the table;

(e) For c-arm type fluoroscopic equipment having a SID less than
forty-five centimeters, the air kerma rate shall be determined at the minimum
SSD; and

(f) The maximum air kerma rate shall be determined with the kVp,
mA and/or other selectable parameters adjusted to those settings which give the
maximum air kerma rate. X-ray systems that incorporate automatic exposure
control shall have sufficient attenuative material placed in the useful beam to
produce the maximum exposure rate of the system.

(D) Handlers of mobile fluoroscopic equipment shall not be
required to comply with the requirements of paragraphs (A)(10), and (A)(11) of
this rule and paragraph (H) of rule 3701:1-66-02 of the Administrative
Code.

(E) Handlers of c-arm fluoroscopic equipment having a maximum SID
less than forty-five centimeters shall not be required to comply with the
requirements of paragraphs (A)(5)(e), (A)(5)(f), (A)(10), (A)(11), and (A)(12)
of this rule and paragraph (H) of rule 3701:1-66-02 of the Administrative Code.
In addition, if a radiation expert has specified in the registrant's
quality assurance program that an individual is unlikely to receive a total
effective dose equivalent of greater than two millirem in any one hour or one
hundred millirem in a year, the handler shall not be required to comply with
the requirements of paragraph (B)(1) of this rule.

(F) All individuals operating fluoroscopic equipment, and
individuals likely to receive an annual effective dose equivalent in excess of
one millisievert (one hundred millirem) from participating in fluoroscopic
procedures, shall receive at least two hours of radiation protection training
specific to fluoroscopy in addition to the training required by rule
3701:1-38-10 of the Administrative Code prior to performing or participating in
fluoroscopic procedures. Additionally, each individual shall receive one hour
of re-training whenever the individual receives in excess of thirty per cent of
the allowable occupational dose measured over one calendar year.

(G) The training required by paragraph (F) of this rule shall be
approved by the registrant's designated radiation expert, and be specific
to the type of fluoroscopic equipment used. Documentation of receiving the
training required by paragraph (F) of this rule shall be retained by the
registrant and be available for review upon inspection. At a minimum, training
topics shall include, but not be limited to:

(1) Principles and
operation of the fluoroscopic equipment to be used;

(2) Fluoroscopic and
radiographic outputs of each mode of operation, including high-level control
options clinically used;

(3) Dose management,
including dose reduction techniques for fluoroscopic equipment;

(4) Safe operating
procedures of each piece of fluoroscopic equipment that may be used by each
individual.

(5) Units of measurement
and dose, including dose-area product values and air kerma;

(6) Radiation protection
methods for patient and staff;

(7) Basic properties of
radiation; and

(8) Biological effects of
radiation.

(H) Fluoroscopic equipment used for radiation therapy procedures
is regulated pursuant to rule 3701:1-67-09 of the Administrative
Code.

(I) Computed tomography scanners equipped with fluoroscopic
capabilities are regulated pursuant to rule 3701:1-66-10 of the Administrative
Code.

View Latest Documents

preview-icon 5 pages

. w wo ey 99 i i gat COURT OF COMMON PLEAS ¢.; f ane BUTLER COUNTY, OHIO ot g 7, vie \ A wie X 20 aw ro Wire, | 3°05 1576 BUTLER cana OF HEALTH | Case No. Lok i \ iv . et - 130 High Sty) buF ; Hamilton, ot eae Oey Plaintiff RTS COMPLAINT FOR INJUNCTION v. JOSEPH D. ELKINS 3900 Withrow Rd. | Hamilton, OH 45011 i Defendant(s) 1. Plaintiff is the board of health of the Butler County General Health District. 2. Defendant(is) is/are the owner(s) of real property locat…

Case Filed

May 24, 2013

County

Butler County, OH

Filed Date

May 24, 2013

Judge Hon. Andrew Nastoff Trellis Spinner 👉 Discover key insights by exploring more analytics for Andrew Nastoff
preview-icon 7 pages

& ~ ed e Ww 98 yA\ COURT OF COMMON PLEAS 4 aoe ees COUNTY, OHIO 2013 05 1572 TLER EALTH | ‘ase No. Meaty BUTLER CO SR | Case N oy eat 301 S. Third St@U ge BY i Judge BUTLER COUNTY Hamilton, OH 4 | CLERK OF COURTS Plaintiff v. | COMPLAINT FOR INJUNCTION WILLIAM JONES JR. i 3206 Trenton Oxford Rd. Hamilton, OH 45011 CYNTHIA J. JONES 3206 Trenton Oxford Rd. Hamilton, OH 45011 Defendant(s) 1. Plaintiff is the board of health of the Butler County General Health District. 2. Defendant…

Case Filed

May 24, 2013

County

Butler County, OH

Filed Date

May 24, 2013

Judge Hon. Keith M. Spaeth Trellis Spinner 👉 Discover key insights by exploring more analytics for Keith M. Spaeth
preview-icon 5 pages

ow w ay tee i, “ 4 Sect y gat 3! COURT OF COMMON PLEAS 20) 2» BUTLER COUNTY, OHIO 3 05 ' ot te ba 975 BUTLER COUNTY Bi SF HEALTH | — Case No. Quibig 2% 301 S. ThinE SKS oF Judge CLERIE Ge COUNTY Hamilton, AY @891 1 | COURTS Plaintiff | v. | COMPLAINT FOR INJUNCTION THELMA HACKER : 7431 Kyles Station Rd. Liberty Township, OH 45044-9424 i Defendant(s) i 1. Plaintiff is the board of health of the Butler County General Health District. 2. Defendant(is) is/are the owner(s) of real property loc…

County

Butler County, OH

Filed Date

May 24, 2013

Category

CIVIL - GENERAL

Judge

Hedric, Honorable Craig D.

preview-icon 4 pages

ELECTRONICALLY FILED COURT OF COMMON PLEAS Monday, February 8, 2021 11:37:00 PM CASE NUMBER: 2020 CR 03129 Docket ID: 35220669 …

Case Filed

Oct 05, 2020

Case Status

CLOSED

County

Montgomery County, OH

Filed Date

Feb 08, 2021

preview-icon 4 pages

E0137 - C60 e ‘ ~ OHIO DEPT. OF HEALTH IN THE COURT OF COMMON PLEAS OF FRANKLIN co! CIVIL DIVISION 2010 JUN 23 mM ea GENERAL COUNSEL STAN'S BAR & GRILL INC. 757 West 2™ Street Xenia, Ohio 45385 Appellant, v carne, JO CVF06 * 9351 OHIO DEPARTMENT OF HEALTH CD ow oo eS 3% Ee * > & YE 246 North High Street es es Columbus, Ohio 43215 - 8S cam : ¢ Be Appellee. S 2 nO =a = =e NOTICE OF APPEAL Ye %B oe < In accordance with the provisions of R.C. 119 12, Appellant hereby gives notice of its a…

Case Name 10 CV 009351
Case Filed

Jun 23, 2010

Case Status

CLOSED

County

Franklin County, OH

Filed Date

Jun 23, 2010

Judge Hon. ANDY D MILLER Trellis Spinner 👉 Discover key insights by exploring more analytics for ANDY D MILLER
View More Documents

Please wait a moment while we load this page.

New Envelope