Ohio Administrative Code|Rule 3701:1-66-06 | Dental radiation-generating equipment.

                                                

(A) Dental equipment shall meet the
following equipment standards:

(1) For dental intraoral
equipment, a means shall be provided to limit the source-to-skin distance (SSD)
to not less than eighteen centimeters;

(2) For dental intraoral
equipment, the x-ray field at the minimum SSD shall be containable in a circle
having a diameter of no more than seven centimeters;

(3) A means shall be
provided to terminate the exposure at a preset time interval, a preset product
of current and time, a preset number of pulses, or a preset radiation exposure
to the image receptor;

(4) The exposure control
switch shall meet the following requirements:

(a) The switch shall be of the "dead-man"
type;

(b) The operator shall be able to terminate the exposure at any
time during an exposure of greater than one-half second. Except during
panoramic radiography, termination of the exposure shall cause automatic
resetting of the timer to its initial setting or to zero; and

(c) It shall not be possible to make an exposure when the timer
is set to "zero" or "off" position if either position is
provided;

(5) The kVp accuracy
shall be within plus or minus ten per cent of the indicated value;

(6) For manual exposures,
the accuracy of the timing device shall be within plus or minus ten per cent of
the indicated setting. The timing device shall be tested at a minimum of two
settings within the operative range of fifty milliseconds to one thousand
milliseconds;

(7) Visual indication
shall be provided whenever x-rays are produced. Certified equipment also shall
provide audible indication to the operator while x-rays are produced or on
termination of the exposure; and

(8) The coefficient of
variation for reproducibility of kVp, timing, and radiation exposure shall not
exceed 0.05 for four consecutive exposures.

(B) In addition to other structural
shielding requirements in rule 3701:1-66-02 of the Administrative Code,
handlers of dental equipment shall comply with the following:

(1) Intraoral and panoral
units shall be provided with primary barriers at all areas struck by the useful
beam. Consideration may be given to the attenuation provided by the patient as
a result of direct interaction with the useful beam;

(2) When intraoral or
panoral units are in adjacent human patient occupied rooms or areas, protective
barriers shall be provided between the rooms or areas, unless safety procedures
are documented and implemented to require that no human patients shall be
present in the adjacent rooms or areas while exposures are being
made;

(3) Intraoral and panoral
units shall be provided with a protective barrier for the operator or shall be
so arranged that the operator is located at a minimum distance of six feet from
the patient and out of the useful beam. The operators position shall be
arranged so that the operator views the patient during the entire exposure;
and

(4) When the operator is
behind a protective barrier, a viewing system shall be provided large enough
and so placed that the operator can see the patient without having to leave the
protected area during exposure.

(C) In addition to the radiation safety
requirements listed in rule 3701:1-66-02 of the Administrative Code, handlers
of dental equipment shall not permit any individual to hold any part of the
x-ray tube housing, cone, or mechanical support of the x-ray tube during
exposure.

(D) In addition to the quality assurance
requirements of rules 3701:1-66-02 and 3701:1-66-04 of the Administrative Code,
handlers of dental radiation-generating equipment shall conduct annual
evaluations of x-ray operators to include the following:

(1) Positioning of the
x-ray tube;

(2) Image
processing;

(3) Operator location
during x-ray exposure;

(4) Appropriate
radiologic protocol; and

(5) applicable regulatory
requirements.

(E) Handlers of dental panoral equipment
shall comply with all requirements of paragraphs (A) to (D) of this rule,
except for paragraphs (A)(1) and (A)(2) of this rule, and must comply with the
following:

(1) Dental panoral x-ray
machines shall be certified pursuant to 21 C.F.R. part 1020 (as effective on
the effective date of this rule); and

(2) The x-ray field shall be limited to
the dimensions of the slit in the image receptor holder or limited to the
dimensions of the active portion of the image receptor.

(F) Except for dental equipment used for
panoral use, handlers of radiographic equipment used for extraoral dental
procedures shall comply with the requirements of paragraphs (A) and (B) of rule
3701:1-66-05 of the Administrative Code.

(G) Fluoroscopy without image
intensification shall not be used for dental examinations. Handlers of image
intensified fluoroscopic equipment shall comply with the applicable
requirements of rule 3701:1-66-07 of the Administrative Code.

(H) Handlers of dental cone-beam CT
radiation-generating equipment shall comply with the applicable requirements of
rule 3701:1-66-10 of the Administrative Code.

(I) Handlers of hand-held
radiation-generating equipment used for dental procedures shall meet the
requirements of paragraphs (A), (B), and (D) of this rule, and shall develop
and implement safe operating procedures as part of the quality assurance
program specified in rule 3701:1-66-04 of the Administrative Code, which shall
address at least the following:

(1) Hand-held
radiation-generating equipment shall be used for intraoral purposes
only;

(2) Operators of the
hand-held radiation-generating equipment shall wear a full lead apron of not
less than 0.25 millimeter lead equivalent;

(3) Hand-held
radiation-generating equipment shall be equipped with a backscatter shield of
not less than 0.25 millimeters lead equivalent and 15.2 centimeters in diameter
that is positioned as close as practicable to the distal end of the
device;

(4) Storage and security procedures shall
be developed and implemented to assure hand-held radiation-generating equipment
is secured against unauthorized use or removal when not under the control and
constant surveillance of the registrant;

(5) Operator training, as required in
paragraph (B)(5) of rule 3701:1-66-04 of the Administrative Code, shall include
documented specific instruction to the x-ray operator regarding the prohibition
on placing any part of their body into the useful beam and ensuring there are
no bystanders within a radius of at least six feet from the patient being
examined during exposure; and

(6) Hand-held
radiation-generating equipment shall not be used in hallways or waiting
rooms.

(J) Dental equipment shall not be operated at less than a
measured fifty-one kVp.

(K) Dental equipment used by veterinarians shall comply with all
requirements of this rule except paragraphs (A)(1), (D), (E), (I) and (J) of
this rule. Additionally, the useful beam shall be limited to the area of
clinical interest.

(L) Extraoral dental equipment used by veterinarians shall follow
the requirements of paragraph (F) of rule 3701:1-66-05 of the Administrative
Code.

View Latest Documents

preview-icon 5 pages

. w wo ey 99 i i gat COURT OF COMMON PLEAS ¢.; f ane BUTLER COUNTY, OHIO ot g 7, vie \ A wie X 20 aw ro Wire, | 3°05 1576 BUTLER cana OF HEALTH | Case No. Lok i \ iv . et - 130 High Sty) buF ; Hamilton, ot eae Oey Plaintiff RTS COMPLAINT FOR INJUNCTION v. JOSEPH D. ELKINS 3900 Withrow Rd. | Hamilton, OH 45011 i Defendant(s) 1. Plaintiff is the board of health of the Butler County General Health District. 2. Defendant(is) is/are the owner(s) of real property locat…

Case Filed

May 24, 2013

County

Butler County, OH

Filed Date

May 24, 2013

Judge Hon. Andrew Nastoff Trellis Spinner 👉 Discover key insights by exploring more analytics for Andrew Nastoff
preview-icon 7 pages

& ~ ed e Ww 98 yA\ COURT OF COMMON PLEAS 4 aoe ees COUNTY, OHIO 2013 05 1572 TLER EALTH | ‘ase No. Meaty BUTLER CO SR | Case N oy eat 301 S. Third St@U ge BY i Judge BUTLER COUNTY Hamilton, OH 4 | CLERK OF COURTS Plaintiff v. | COMPLAINT FOR INJUNCTION WILLIAM JONES JR. i 3206 Trenton Oxford Rd. Hamilton, OH 45011 CYNTHIA J. JONES 3206 Trenton Oxford Rd. Hamilton, OH 45011 Defendant(s) 1. Plaintiff is the board of health of the Butler County General Health District. 2. Defendant…

Case Filed

May 24, 2013

County

Butler County, OH

Filed Date

May 24, 2013

Judge Hon. Keith M. Spaeth Trellis Spinner 👉 Discover key insights by exploring more analytics for Keith M. Spaeth
preview-icon 5 pages

ow w ay tee i, “ 4 Sect y gat 3! COURT OF COMMON PLEAS 20) 2» BUTLER COUNTY, OHIO 3 05 ' ot te ba 975 BUTLER COUNTY Bi SF HEALTH | — Case No. Quibig 2% 301 S. ThinE SKS oF Judge CLERIE Ge COUNTY Hamilton, AY @891 1 | COURTS Plaintiff | v. | COMPLAINT FOR INJUNCTION THELMA HACKER : 7431 Kyles Station Rd. Liberty Township, OH 45044-9424 i Defendant(s) i 1. Plaintiff is the board of health of the Butler County General Health District. 2. Defendant(is) is/are the owner(s) of real property loc…

County

Butler County, OH

Filed Date

May 24, 2013

Category

CIVIL - GENERAL

Judge

Hedric, Honorable Craig D.

preview-icon 4 pages

ELECTRONICALLY FILED COURT OF COMMON PLEAS Monday, February 8, 2021 11:37:00 PM CASE NUMBER: 2020 CR 03129 Docket ID: 35220669 …

Case Filed

Oct 05, 2020

Case Status

CLOSED

County

Montgomery County, OH

Filed Date

Feb 08, 2021

preview-icon 4 pages

E0137 - C60 e ‘ ~ OHIO DEPT. OF HEALTH IN THE COURT OF COMMON PLEAS OF FRANKLIN co! CIVIL DIVISION 2010 JUN 23 mM ea GENERAL COUNSEL STAN'S BAR & GRILL INC. 757 West 2™ Street Xenia, Ohio 45385 Appellant, v carne, JO CVF06 * 9351 OHIO DEPARTMENT OF HEALTH CD ow oo eS 3% Ee * > & YE 246 North High Street es es Columbus, Ohio 43215 - 8S cam : ¢ Be Appellee. S 2 nO =a = =e NOTICE OF APPEAL Ye %B oe < In accordance with the provisions of R.C. 119 12, Appellant hereby gives notice of its a…

Case Name 10 CV 009351
Case Filed

Jun 23, 2010

Case Status

CLOSED

County

Franklin County, OH

Filed Date

Jun 23, 2010

Judge Hon. ANDY D MILLER Trellis Spinner 👉 Discover key insights by exploring more analytics for ANDY D MILLER
View More Documents

Please wait a moment while we load this page.

New Envelope