Ohio Administrative Code|Rule 3701:1-67-09 | Quality assurance for radiation therapy , simulation and image guidance equipment.

                                                

(A) For therapy equipment subject to this
chapter, a qualified medical physicist shall develop a documented quality
assurance (QA) program using the appropriate "American Association of
Physicists in Medicine" (AAPM) reports or the "National Council of
Radiation Protection" (NCRP) report as a basis.

(1) The quality assurance
program shall:

(a) Identify each QA performance test to be
performed;

(b) Describe the procedures used to complete each QA performance
test;

(c) Describe the method used to document the results of each QA
performance test;

(d) Identify the frequency of each QA performance test;
and

(e) Specify the acceptable action limits and safety tolerance
limits for each QA performance test result and the action to be taken when
exceeded.

(2) Any variation from
the identified tests, frequency or tolerance limits specified in the
appropriate AAPM or NCRP reports shall be based on a documented history of
therapy equipment performance or inherent therapy equipment design and be
justified in the quality assurance program.

(3) Any QA performance
test result exceeding a factor of two from the tolerances in AAPM or NCRP
documents shall require immediate action prior to further
treatment.

(B) The handler shall perform QA
performance tests in accordance with the written procedures established by the
qualified medical physicist and comply with the following:

(1) The authorized user
and qualified medical physicist shall be immediately notified if any QA
performance test result exceeds a safety tolerance limit set by the qualified
medical physicist. The cause for a parameter exceeding the safety tolerance
limit shall be investigated before the system is used for patient irradiation.
The medical physicist in collaboration with the authorized user shall determine
whether medical treatment may continue safely or be interrupted until
corrected;

(2) The handler shall use
a dosimetry system described in rule 3701:1-67-07 of the Administrative Code to
perform absolute-dose related QA performance tests required by this
rule;

(3) The handler shall
have the qualified medical physicist review and sign the results of each QA
performance test within a month of the date that the test was performed, by an
individual other than the qualified medical physicist;

(4) The handler shall
ensure that safety QA tests are performed monthly on the
following:

(a) Electrical interlocks at each external beam radiation therapy
room entrance;

(b) The "beam-on" and termination
switches;

(c) All beam indicator lights;

(d) Patient audio visual viewing system; and

(e) If applicable, electrically operated treatment room doors
from inside and outside the treatment room.

(C) As used in this rule,
"calibration" means the determination of the exposure or dose per
unit time or absorbed dose per monitor unit (MU) under specified conditions as
described in the quality assurance program. Calibration shall be
performed:

(1) Before the first
medical use following installation or reinstallation;

(2) Annually;
and

(3) Before medical use
under the following conditions:

(a) Whenever QA performance test results indicate the radiation
output differs by more than five per cent from the calibration value obtained
during the most recent annual QA performance tests and the difference cannot be
reconciled. Calibration of therapy equipment with multi-energy capabilities is
required only for those modes and/or energies that are not within their
acceptable range; and

(b) Following any major mechanical, electrical or software based
alterations affecting the radiation source, its housing, power supply or
controls or after replacement of the radiation source. If an alteration or
replacement does not affect all energies, calibration shall be performed on the
affected energy that is in most frequent clinical use at the facility. The
remaining energies may be validated with quality assurance check procedures
against the criteria in paragraph (C)(3)(a) of this rule.

(D) For therapy equipment operating at
less than one megavolt (MV):

(1) The qualified medical
physicist shall use NCRP report 69, Dosimetry of X-Ray and Gamma Ray Beams for
Radiation Therapy in the Energy Range 10 keV to 50MeV (1981) for commissioning,
initial QA performance testing and to meet the requirements of paragraph (A) of
this rule. The term "QA performance test", as used in paragraph (D)
of this rule, shall have the same meaning as the term "check" in
the NCRP report.

(2) Commissioning and
initial QA performance testing shall be completed prior to medical use
following installation or reinstallation.

(3) Commissioning,
initial and annual QA performance tests shall be performed by or under direct
supervision of a qualified medical physicist.

(E) For therapy equipment operating at or
above one megavolt (MV):

(1) The qualified medical
physicist shall use the "AAPM Code of Practice for Radiotherapy
Accelerators: AAPM Report No. 47 (AAPM report 47)," prepared by
"Radiation Therapy Task Group 45" (this publication can be obtained
from the American association of physicists in medicine, One Physics Ellipse,
College Park, MD 20740, telephone (301) 209-3350,
http://www.aapm.org/pubs/reports) and the manufacturer's contractual
specifications as a basis for acceptance testing and
commissioning.

(2) To meet the
requirements of paragraph (A) of this rule, the qualified medical physicist
shall use:

(a) The "Comprehensive QA for Radiation Oncology:
Report of AAPM Radiation Therapy Committee Task Group No. 40: AAPM Report No.
46 (AAPM report 46)" (this publication can be obtained from the American
association of physicists in medicine, One Physics Ellipse, College Park, MD
20740, telephone (301) 209-3350,
http://www.aapm.org/pubs/reports).

(b) "Task Group 142 report: Quality assurance of
medical accelerators" (this publication can be obtained from the American
association of physicists in medicine, One Physics Ellipse, College Park, MD
20740, telephone (301) 209-3350, http://www.aapm.org/pubs/reports) for therapy
equipment provided with asymmetric jaws, multileaf collimation, dynamic or
virtual wedges, planar imaging devices, tomographic imaging devices or those
used for stereotactic radiosurgery, stereotactic body radiation therapy, total
body photon irradiation or intensity-modulated radiotherapy.

(c) The "Intraoperative radiation therapy using mobile
electron linear accelerators: Report of AAPM Radiation Therapy Committee Task
Group No. 72" (this publication can be obtained from the American
association of physicists in medicine, One Physics Ellipse, College Park, MD
20740, telephone (301) 209-3350, http://www.aapm.org/pubs/reports) for mobile
electron linear accelerator therapy equipment.

(d) The "Report of AAPM TG 135: Quality assurance for
robotic radiosurgery" (this publication can be obtained from the American
association of physicists in medicine, One Physics Ellipse, College Park, MD
20740, telephone (301) 209-3350, http://www.aapm.org/pubs/reports) for robotic
radiosurgery therapy equipment.

(e) The "QA for helical tomotherapy: Report of the
AAPM Task Group 148" (this publication can be obtained from the American
association of physicists in medicine, One Physics Ellipse, College Park, MD
20740, telephone (301) 209-3350, http://www.aapm.org/pubs/reports) for helical
tomotherapy equipment.

(3) Acceptance testing, commissioning and
baseline QA performance testing shall be completed prior to medical use
following installation or reinstallation.

(4) Acceptance testing, commissioning,
baseline and annual QA performance tests shall be performed by or under direct
supervision of a qualified medical physicist.

(5) An independent verification of the
calibration of all photon beams and a sample of available electron beams shall
be performed annually by:

(a) A second radiation expert using a dosimetry system other than
the dosimetry system that was used during the annual calibration;
or

(b) A national institute of science and technology traceable
third-party dosimetry service or an equivalent method which is capable of
measuring doses with an accuracy within five percent.

(6) Proper operation of each emergency
power cutoff switch shall be verified annually. If more than one switch is
installed, they may be evaluated on a rotating basis throughout the
year.

(F) For conventional or virtual
simulation:

(1) The qualified medical
physicist shall use the "Comprehensive QA for Radiation Oncology: Report
of AAPM Radiation Therapy Committee Task Group No. 40: AAPM Report No. 46 (AAPM
report 46)" (this publication can be obtained from the American
association of physicists in medicine, One Physics Ellipse, College Park, MD
20740, telephone (301) 209-3350, http://www.aapm.org/pubs/reports) to meet the
requirements of paragraph (A) of this rule for a conventional simulator;
or

(2) The qualified
medical physicist shall use the "Quality assurance for computed tomography
simulators and the computed tomography-simulation process: Report of the AAPM
Radiation Therapy Committee Task Group No. 66: AAPM Report No. 83 (AAPM report
83)" (this publication can be obtained from the American association of
physicists in medicine, One Physics Ellipse, College Park, MD 20740, telephone
(301) 209-3350, http://www.aapm.org/pubs/reports) for acceptance testing,
commissioning and to meet the requirements of paragraph (A) of this rule for a
virtual simulator.

(3) Acceptance testing,
commissioning, initial QA performance testing, annual QA performance testing
and semiannual (if appropriate) QA performance testing shall be performed by or
under the direct supervision of a qualified medical physicist.

(G) For therapy equipment used for IMRT,
patient specific treatment QC shall be performed before the first fraction is
delivered unless extenuating circumstances are documented by the medical
physicist. If a direct measurement for individual plans is not performed, the
checks shall include both a dose calculation second check and a method to
validate patient plan transfer and deliverability to the treatment
unit.

(H) The handler shall maintain a record
of each QA performance test result for three years and each calibration for the
duration of the registration. The records shall include:

(1) The date of the QA
test or calibration;

(2) The
manufacturer's name, model number, and serial number of the therapy
equipment;

(3) The
manufacturer's name, model numbers and serial numbers for the
instrument(s) used to measure the radiation output of the therapy equipment;
and

(4) The signature of the
individual who performed the QA performance test or calibration.

View Latest Documents

preview-icon 5 pages

. w wo ey 99 i i gat COURT OF COMMON PLEAS ¢.; f ane BUTLER COUNTY, OHIO ot g 7, vie \ A wie X 20 aw ro Wire, | 3°05 1576 BUTLER cana OF HEALTH | Case No. Lok i \ iv . et - 130 High Sty) buF ; Hamilton, ot eae Oey Plaintiff RTS COMPLAINT FOR INJUNCTION v. JOSEPH D. ELKINS 3900 Withrow Rd. | Hamilton, OH 45011 i Defendant(s) 1. Plaintiff is the board of health of the Butler County General Health District. 2. Defendant(is) is/are the owner(s) of real property locat…

Case Filed

May 24, 2013

County

Butler County, OH

Filed Date

May 24, 2013

Judge Hon. Andrew Nastoff Trellis Spinner 👉 Discover key insights by exploring more analytics for Andrew Nastoff
preview-icon 7 pages

& ~ ed e Ww 98 yA\ COURT OF COMMON PLEAS 4 aoe ees COUNTY, OHIO 2013 05 1572 TLER EALTH | ‘ase No. Meaty BUTLER CO SR | Case N oy eat 301 S. Third St@U ge BY i Judge BUTLER COUNTY Hamilton, OH 4 | CLERK OF COURTS Plaintiff v. | COMPLAINT FOR INJUNCTION WILLIAM JONES JR. i 3206 Trenton Oxford Rd. Hamilton, OH 45011 CYNTHIA J. JONES 3206 Trenton Oxford Rd. Hamilton, OH 45011 Defendant(s) 1. Plaintiff is the board of health of the Butler County General Health District. 2. Defendant…

Case Filed

May 24, 2013

County

Butler County, OH

Filed Date

May 24, 2013

Judge Hon. Keith M. Spaeth Trellis Spinner 👉 Discover key insights by exploring more analytics for Keith M. Spaeth
preview-icon 5 pages

ow w ay tee i, “ 4 Sect y gat 3! COURT OF COMMON PLEAS 20) 2» BUTLER COUNTY, OHIO 3 05 ' ot te ba 975 BUTLER COUNTY Bi SF HEALTH | — Case No. Quibig 2% 301 S. ThinE SKS oF Judge CLERIE Ge COUNTY Hamilton, AY @891 1 | COURTS Plaintiff | v. | COMPLAINT FOR INJUNCTION THELMA HACKER : 7431 Kyles Station Rd. Liberty Township, OH 45044-9424 i Defendant(s) i 1. Plaintiff is the board of health of the Butler County General Health District. 2. Defendant(is) is/are the owner(s) of real property loc…

County

Butler County, OH

Filed Date

May 24, 2013

Category

CIVIL - GENERAL

Judge

Hedric, Honorable Craig D.

preview-icon 4 pages

ELECTRONICALLY FILED COURT OF COMMON PLEAS Monday, February 8, 2021 11:37:00 PM CASE NUMBER: 2020 CR 03129 Docket ID: 35220669 …

Case Filed

Oct 05, 2020

Case Status

CLOSED

County

Montgomery County, OH

Filed Date

Feb 08, 2021

preview-icon 4 pages

E0137 - C60 e ‘ ~ OHIO DEPT. OF HEALTH IN THE COURT OF COMMON PLEAS OF FRANKLIN co! CIVIL DIVISION 2010 JUN 23 mM ea GENERAL COUNSEL STAN'S BAR & GRILL INC. 757 West 2™ Street Xenia, Ohio 45385 Appellant, v carne, JO CVF06 * 9351 OHIO DEPARTMENT OF HEALTH CD ow oo eS 3% Ee * > & YE 246 North High Street es es Columbus, Ohio 43215 - 8S cam : ¢ Be Appellee. S 2 nO =a = =e NOTICE OF APPEAL Ye %B oe < In accordance with the provisions of R.C. 119 12, Appellant hereby gives notice of its a…

Case Name 10 CV 009351
Case Filed

Jun 23, 2010

Case Status

CLOSED

County

Franklin County, OH

Filed Date

Jun 23, 2010

Judge Hon. ANDY D MILLER Trellis Spinner 👉 Discover key insights by exploring more analytics for ANDY D MILLER
View More Documents

Please wait a moment while we load this page.

New Envelope